Performance Tests 1
Keysight 53210A/53220A/53230A Assembly Level Service Guide 77
Test 7a: (Q) Pulse burst frequency - 53230A channel 3 option 150 with
option 106
This test uses the 53230A to measure both narrow and wide pulse outputs of a
6 GHz carrier.
Equipment
– FS725 Rubidium timebase
– Keysight E8257D-520-UNW microwave signal generator
– Keysight 10503A BNC cable (2)
– Keysight 11500E (3.5) mm cable
– Keysight 1250-1250 N(M) to SMA(F)
E8257D-520-UNW setup
– Output Frequency: 6 GHz
– Output Power: 0 dBm
–Auto-Level Control: Off
–Pulse Mode: On
– Pulse Repetition Interval Period: 1 mSec
– Pulse Width (Narrow): 200 nSec
NOTE
This test procedure applies only to the 53230A Counter with Option 106 and
Option 150 installed. Option 106 is installed as microwave channel 3.
NOTE
The Quick Verification (Q) requires only the Narrow Pulse Width Burst Frequency