Keysight 53210A/53220A/53230A Assembly Level Service Guide 179
+907, “Self-test failed; PIC communication failure”
(1) 53200-66501 Motherboard; (2) Battery; (3) P0500-63001 processor
board
+908, “Self-test failed; Battery Failed”
(1) 1420-0909 battery; (2) 53200-66507 battery charger assy
+909, “Self-test failed; GPIB test Failed”
(1) P0001-20001 GPIB board; (2) P0500-63001 processor board
+910, “Self-test failed; Channel 3 test failed”
(1) 53200-66505 (Option 106) or 53200-66506 (Option 115);
(2) 532xx-66501 motherboard
+911, “Self-test failed; Front Panel revision check failed”
(1) 53200-66502 front panel; (2) P0500-63001 processor board
+912, “Self-test failed; Measurement board test failed”
(1) 53200-66501 Motherboard; (2) P0500-63001 processor board