6Backdating
270 Keysight 53210A/53220A/53230A Assembly Level Service Guide
For the 53220A or 53230A Counters, send the command to calibrate as:
“CALibration:LEVel3? 500E8, -27”.
For the 53210A Counter, send the command to calibrate as:
“CALibration:LEVel2? 500E8, -27”.
Retrieve query result:
0 indicates the calibration PASSED.
Any value other than 0 indicates calibration FAILED.
If the calibration passes at 500 MHz, repeat the calibration sequence for 1000
MHz at -27 dBm as follows:
For the 53220A or 53230A Counters, send the command to calibrate as:
“CALibration:LEVel3? 1E9, -27”.
For the 53210A Counter, send the command to calibrate as:
“CALibration:LEVel2? 1E9, -27”.
Retrieve query result:
0 indicates the calibration PASSED.
Any value other than 0 indicates calibration FAILED.
If non-zero value for either calibration point, read error queue as follows:
SYSTem:ERRor?
Read errors until no more errors.
–Consult the “Error Messages” on page 174 of Chapter 2 for what to do after
receiving a specific calibration error message.

Table of Contents

Other manuals for Keysight Technologies 53210A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies 53210A and is the answer not in the manual?

Keysight Technologies 53210A Specifications

General IconGeneral
BrandKeysight Technologies
Model53210A
CategoryCash Counter
LanguageEnglish

Related product manuals