Performance Tests 1
Keysight 53210A/53220A/53230A Assembly Level Service Guide 51
4 Press the Single Period softkey.
– The following softkeys are displayed:
5 Press and set:
–Auto Level: Off
– Level: 0 V
–Noise Rej: Off
6 Press the Channel 1 key.
– The Channel 1 Setup menu softkeys are displayed.
7 Press and set:
– Coupling: DC
– Impedance: 50 Ω
– Range: 5 V
–BW Limit: Off
8 Press the Trigger key.
– The trigger softkeys are displayed: