1 Performance Tests
70 Keysight 53210A/53220A/53230A Assembly Level Service Guide
7 Press the Trigger key.
– The counter takes 100 readings and fills in the math functions.
8 Verify the ‘Mean’ is:
– 5.88 GHz ± 58.800 Hz (53210A)
– 5.88 GHz ± 5.880 Hz (53220A/53230A)
9 Verify the ‘Std Dev’ is less than:
– 82.300 Hz (53210A)
– 41.100 Hz (53220A)
– 8.230 Hz (53230A)
10 Record the Mean and Standard Deviation for 5.88 GHz, Test 6a,
in the Keysight 53210A/53220A/53230A Performance Test Record.
11 Press the Reset Stats softkey.
12 If Opt 150 is installed, proceed to Test 7a. Otherwise, disconnect the test
setup.