1 Performance Tests
86 Keysight 53210A/53220A/53230A Assembly Level Service Guide
9 Press the Trigger key.
–The Trigger softkeys are displayed:
10 Press to set:
–Source: Manual
– Wait For Trig is displayed
–Rdgs/Trigger: 100
–Trig Delay: 0 seconds
11 Press the Math key.
–The Math softkeys are displayed:
12 Press to set:
–All Math: On
13 Press the Statistics softkey:
–The Statistics softkeys are displayed: