Keysight B2900 User’s Guide, Edition 5 2-15
Introduction
Source/Measure Unit
Table 2-6 Voltage Measurement Range
Table 2-7 Resistance Measurement Range
1
Range value Voltage measurement value Resolution
0.2 V 0 |V| 0.212 V 0.1 V
2 V 0 |V| 2.12 V 1 V
20 V 0 |V| 21.2 V 10 V
200 V 0 |V| 212 V 100 V
Range value Resistance measurement value
Display
resolution
Test
current
2 0 < R 2 1 1 A
20 2 < R 20 10 100 mA
200 20 < R 200 100 10 mA
2 k 200 < R 2 k 1 m 1 mA
20 k 2 k < R 20 k 10 m 100 A
200 k 20 k < R 200 k 100 m 10
A
2 M 200 k < R 2 M 1 1 A
20 M 2 M < R 20 M 10 100 nA
200 M 20 M < R 200 M 100 10 nA
1. The resistance measurement range is effective for the resistance measurements set to the AUTO
mode which is selected by the :SENS:RES:MODE command.