EasyManuals Logo

Keysight Technologies B2900 Series User Manual

Keysight Technologies B2900 Series
234 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #54 background imageLoading...
Page #54 background image
2-14 Keysight B2900 User’s Guide, Edition 5
Introduction
Source/Measure Unit
Table 2-4 Limitations for using Channels 1 and 2
Table 2-5 Current Measurement Range
Channel 1 voltage V1 Channel 2 voltage V2
Current limit
a
a. I1: Channel 1 current, I2: Channel 2 current
0 < |V1| 6 V 0 < |V2| 6 V I1 + I2 4 A
6 V < |V2| 21 V I1 + I2 1.6 4 A
6 V < |V1| 21 V 0 < |V2| 6 V I1 + I2 0.625 2.5 A
6 V < |V2| 21 V I1 + I2 2.5 A
Range value Current measurement value Resolution
10 nA
a
a. Available for the B2911A and B2912A. Not available for the B2901A
and B2902A.
0 |I| 10.6 nA 10 fA
100 nA 0 |I| 106 nA 100 fA
1 A0 |I| 1.06 A1 pA
10 A0 |I| 10.6 A 10 pA
100 A0 |I| 106 A 100 pA
1 mA 0 |I| 1.06 mA 1 nA
10 mA 0 |I| 10.6 mA 10 nA
100 mA 0 |I| 106 mA 100 nA
1 A 0 |I| 1.06 A 1 A
1.5 A 0 |I| 1.53 A
3 A 0 |I| 3.06 A 10 A
10 A
b
b. Available for pulse mode. Not available for DC mode.
0 |I| 10.6 A

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies B2900 Series and is the answer not in the manual?

Keysight Technologies B2900 Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelB2900 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals