CALCulate:LIMit:{LOWer|UPPer}[:DATA] {<value>|MIN|MAX|DEF}
CALCulate:LIMit:{LOWer|UPPer}[:DATA]? [{MIN|MAX|DEF}]
Sets an upper or lower limit.
Parameter Typical Return
-1.0E+15 to -1.0E-15,
or
0.0 (default)
or
+1.0E-15 to 1.0E+15
+1.00000000E+00
See
Example
.
l You can assign a lower limit, an upper limit or both. Do not set the lower limit above the upper limit.
l
Limit crossing: If a measurement is less than the specified lower limit, bit 11 ("Lower Limit Failed") is
set in the Questionable Data Condition Register. A measurement greater than the specified upper limit
sets bit 12 ("Upper Limit Failed"). See STATus Subsystem Introduction for further information.
l CONFigure resets both limits to 0.
l This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset
(SYSTem:PRESet).