CALCulate:LIMit[:STATe] {ON|1|OFF|0}
CALCulate:LIMit[:STATe]?
Enables or disables limit testing.
Parameter Typical Return
{ON|1|OFF|0}. Default: OFF. 0(OFF)or1(ON)
See
Example
.
l The instrument clears front panel indications of limits being exceeded and clears bits 11 and 12 in the
Questionable Data Register when the measurement function changes, or when any of the following
commands is executed:
CALCulate:LIMit:STATe ON
INITiate
MEASure:<function>?
READ?
CALCulate:LIMit:CLEar
*RST
SYSTem:PRESet
l The instrument turns this setting OFF when the measurement function is changed or after *RST or
SYSTem:PRESet.
SCPI Programming Reference
216
Keysight Truevolt Series Operating and Service Guide