CALCulate:LIMit:CLEar[:IMMediate]
Clears front panel indications of limits being exceeded and clears bit 11 ("Lower Limit Failed") and bit 12
("Upper Limit Failed") in the Condition Register of the Questionable Data Register event register group.
The corresponding event register bits are unaffected.
A condition register continuously monitors the state of the instrument. Condition register bits are updated
in real time; they are neither latched nor buffered.
An event register is a read-only register that latches events from the condition register. While an event bit
is set, subsequent events corresponding to that bit are ignored.
Parameter Typical Return
(none) (none)
Clear the limit test results:
CALC:LIM:CLE
l Does not clear measurements in reading memory.
l The instrument clears front panel indications of limits being exceeded and clears bits 11 and 12 in the
Questionable Data Register when the measurement function changes, or when any of the following
commands is executed:
CALCulate:LIMit:STATe ON
INITiate
MEASure:<function>?
READ?
CALCulate:LIMit:CLEar
*RST
SYSTem:PRESet
l To clear statistics, limits, histogram data, and measurement data, use CALCulate:CLEar[:IMMediate].
SCPI Programming Reference
214
Keysight Truevolt Series Operating and Service Guide

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Truevolt Series and is the answer not in the manual?

Keysight Truevolt Series Specifications

General IconGeneral
BrandKeysight
ModelTruevolt Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals