Medrad™ Intego™
PET Infusion System
90
4. The Intego™ PET Infusion System displays a dialog box prompting the clinician to insert
the
57
Co source into the Dose Calibrator. Using the Calibration Source Holder, insert the
57
Co source into the Dose Calibrator and close the Shielded Chamber Lid.
5. Confirm the assay information listed for the first source is correct. If it is not, press the
EDIT button to make changes. Once the information is correct, press the OK button.