A7™ Operating Instructions 046-004667-00 4 - 17
Preoperative Tests Automatic Backup Flow Control Test
3. Compare the test results with the information in TABLE 4-5, “Automatic Backup Flow Control
Test,” on page 4-17, and proceed accordingly.
RESULTS COMMENTS/OPTIONS
Pass
Example:
Automatic Backup Flow Control Test passed.
Select Continue to proceed to the main screen.
Fail
Example:
Automatic Backup Flow Control Test failed.
Select Retry to repeat the test.
or
Select Accept to proceed without Automatic Backup Flow
Control deployment.
TABLE 4-5 Automatic Backup Flow Control Test