A7™ Operating Instructions 046-004667-00 3 - 13
System Interface Fresh Gas Flow Display
NOTE: OPTIMIZER® is active only when all the following conditions are
met.
EFCS is active.
External AG module is active.
The anesthesia system is a in mechanical ventilation mode.
NOTE: OPTIMIZER® is disabled if automatic circuit leak test is skipped or
failed.
NOTE:
OPTIMIZER® is disabled if the measurement value of Vt, MV,
EtCO2, FiCO2, or Rate is invalid.
NOTE: If fresh gas data is unavailable, the
OPTIMIZER® will be
ineffective.
NOTE:
OPTIMIZER® is disabled if any of the following alarms appears:
•Apnea
Apnea > 2 min
Apnea CO2
•Flow Sensor Failure
•Check Flow Sensors
Pinsp Not Achieved
Vt Not Achieved
Patient Circuit Leak
CO2 Absorber Canister not Locked
Ventilator Comm Stop
Drive Gas Pressure Low
AG Hardware Error
External AG Self Test Error
AG Hardware Malfunction
AG Init Error
•AG No Watertrap
•AG Change Watertrap
•AG Comm Stop
•AG Airway Occluded
•AG Zero Failed
External AG Module Disconnected
Incompatible AG Software Version
FIGURE 3-20 Inactive
OPTIMIZER®
NOTE: If fresh gas data is unavailable, the
OPTIMIZER® will be
ineffective.
Inactive OPTIMIZER®

Table of Contents

Other manuals for Mindray A7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A7 and is the answer not in the manual?

Mindray A7 Specifications

General IconGeneral
BrandMindray
ModelA7
CategoryMedical Equipment
LanguageEnglish

Related product manuals