6-35
6. Select Cancel. The screen shown below is displayed.
7. The screen shown below is displayed after the first step of manual calibration is completed.
The system calibrates 16 calibration points one by one. All 16 points are on the rising curve.
You can select to recalibrate any calibration point during calibration. When a calibration
point mapped to the falling curve is calibrated, you cannot change the points mapping to the
rising curve. If Waiting is displayed in the cell, wait for the system to implement pressure
control. If Input Cal Value is displayed in the cell, enter the standard pressure displayed on
the calibration device. After entering standard pressures of all the 16 calibration points,
select Done to check and save the calibration data.

Table of Contents

Other manuals for Mindray A9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A9 and is the answer not in the manual?

Mindray A9 Specifications

General IconGeneral
TypeAnesthesia Workstation
Integrated Gas AnalysisYes
Backup BatteryYes
Operating Temperature10°C to 40°C
Storage Temperature-20°C to 55°C
Battery TypeLithium-ion
Ventilation ModesVolume Control, Pressure Control, SIMV, Manual Ventilation
DisplayTouchscreen
Patient MonitoringECG, SpO2, NIBP, EtCO2, Temperature
ConnectivityEthernet, USB
PortabilityMobile (with wheels)
Battery LifeMinimum 90 minutes (typical)
AltitudeUp to 3000 meters
Power Supply100-240 VAC, 50/60 Hz
Charging TimeApproximately 4 hours
StorageInternal storage for patient data and system logs
Supported LanguagesMultiple languages
CertificationsCE, FDA