6-37 
 
9.  The screen shown below is displayed if the calibration fails. 
 
 
6.3.5.4 Common Failures and Recommended Actions 
Failure 
Description 
Possible Cause  Recommended Action 
Display 1 
An error occurs in the communication 
with the calibration device. 
1.  Restart the anesthesia machine 
and/or calibration device.   
2.  Check the communication 
cable. 
3.  Replace the VCM or 
calibration device. 
Display 2  The zero point of the calibration device 
is beyond the valid range of [–0.3, 0.3] 
cmH2O. 
Zero the calibration device. 
Display 3 
The pressure cannot be controlled 
within the range of [1, 3] cmH2O. 
1.  Use the valve diagnosis tool 
for diagnosis. Open the PEEP 
valve within the range of 
0–500 mA. If the AD values 
collected by the inspiratory 
pressure sensor increase 
gradually, recalibration is 
recommended.