6-60
Failure
Description
Possible Cause Recommended Action
2. Check whether the electronic
flowmeter works properly.
00000008
The temperature does not meet the
requirement.
1. Replace the electronic
vaporizer.
00000010 The AG module is not ready. 1. Check whether the AG
module is connected and
works in full accuracy mode.
2. Check whether the water trap
and sampling line of the AG
module are connected
properly and whether leakage
occurs.
3. Check whether there is any
other type of anesthetic gas
remaining in the line. If yes,
turn on the fresh gas for
flushing or turn on the injector
of the electronic vaporizer to
be calibrated for flushing.
4. Check whether leakage occurs
in the internal pipeline of the
machine so that the fresh gas
cannot flow through the
sampling point.
5. Replace the internal AG
module.
6. Replace the electronic
vaporizer.
00000020 There is no concentration output. 1. Check whether the AG
module is connected and
works in full accuracy mode.
2. Check whether the water trap
and sampling line of the AG
module are connected
properly and whether leakage
occurs.
3. Check whether there is any
other type of anesthetic gas
remaining in the line. If yes,
turn on the fresh gas for
flushing or turn on the injector
of the electronic vaporizer to
be calibrated for flushing.
4. Check whether leakage occurs
in the internal pipeline of the
machine so that the fresh gas