9 Troubleshooting                                                                                    9-31 
10070001CLM5 
IISE  unit  result  error: 
Clean cycle error 
No ISE wash solution  Refill ISE wash solution 
10070001COM5
ISE  unit  result  error: 
Communication error 
ISE connection error  Reconnect; reset ISE; turn on the power of ISE module .
10070001GAA5 
ISE  unit  result  error: 
Purge  A  cycle error.  Air 
in calibrant A 
Calibrator A bottle empty;   
tube not connected; 
pump A not working; 
tube  blocked,  broken  or 
twisted; 
fibrin and salt in the tube; 
air bubble detector error; 
waste pump error 
Wrong electrode installation. Check the pressing plate 
spring and the sealing to ensure all the electrodes and 
the O-ring are in the right place. 
Wash the module by using <CLEN > program. Remove 
the module and then wash it. Reinstall the sensor. 
Replace air bubble detection sensor;   
replace the waste pump; 
replace the calibrator A and recalibrate; 
reconnect or replace the tube; 
check the connection of the motor; 
replace the pump box; replace the motor; 
replace the tube. 
10070001GAF5 
ISE  unit  result  error: 
Purge A  cycle error. No 
flow 
Calibrator  A  bottle  empty; 
wrong electrode installation 
Replace  the  calibrator  A,  then  recalibrate;  check  the 
installation of the electrodes.