9 Troubleshooting
9-32
10070001GBB5
ISE unit result error:
Purge B cycle error. Air
in calibrant B
Calibrator B bottle empty;
tube not connected;
pump A not working;
tube blocked, broken or
twisted;
fibrin and salt in the tube;
air bubble detector error;
waste pump error
Wrong electrode installation. Check the pressing plate
spring and the sealing to ensure all the electrodes and
the O-ring are in the right place.
Wash the module by using <CLEN > program. Remove
the module and then wash it. Reinstall the sensor.
Replace air bubble detection sensor;
replace the waste pump;
replace the calibrator B and recalibrate;
reconnect or replace the tube;
check the connection of the motor;
replace the pump box; replace the motor;
replace the tube.
10070001GBF5
ISE unit result error:
Purge B cycle error. No
flow
Calibrator B bottle empty;
wrong electrode installation
Replace the calibrator B, then recalibrate; check the
installation of the electrodes.