9 Troubleshooting   
9-46 
100700070045 
ISE  unit  result  error:  K 
electrode  out  of  slope 
range 
Wrong electrode installation; 
calibrator invalidated; 
Electrodes  performance 
degrading. 
air bubble in the electrode; 
reference electrode damaged;
electrodes  interact  with  each 
other; 
module  or  fluid  path 
temperature exceeding  ℃; 
Something  wrong  with  the 
sample 
Reinstall the sensor; 
Rerun after replacing the calibrator B; If the result is still 
abnormal,, rerun after replacing calibrator A; 
rerun after replacing problematic sensors. 
Reinstall and recalibrate, after removing the bubble in 
the electrode; 
rerun after replacing the reference electrode; 
rerun after replacing the K electrode; 
monitor the temperature. If the ambient temperature is 
too high, change the location of the system. 
100700070065 
ISE  unit  result  error:  K, 
Na  electrodes  out  of 
slope range 
Wrong electrode installation; 
calibrator invalidated; 
Electrodes  performance 
degrading. 
air bubble in the electrode; 
reference electrode damaged;
electrodes  interact  with  each 
other; 
module  or  fluid  path 
temperature exceeding  ℃; 
Something  wrong  with  the 
sample 
Reinstall the sensor; 
Rerun after replacing the calibrator B; If the result is still 
abnormal,, rerun after replacing calibrator A; 
rerun after replacing problematic sensors. 
Reinstall and recalibrate, after removing the bubble in 
the electrode; 
rerun after replacing the reference electrode; 
rerun after replacing the K, Na electrode; 
monitor the temperature. If the ambient temperature is 
too high, change the location of the system.