9 Troubleshooting 9-47
100700070085
ISE unit result error: Cl
electrode out of slope
range
Wrong electrode installation;
calibrator invalidated;
Electrodes performance
degrading.
air bubble in the electrode;
reference electrode damaged;
electrodes interact with each
other;
module or fluid path
temperature exceeding ℃;
Something wrong with the
sample
Reinstall the sensor;
Rerun after replacing the calibrator B; If the result is still
abnormal,, rerun after replacing calibrator A;
rerun after replacing problematic sensors.
Reinstall and recalibrate, after removing the bubble in
the electrode;
rerun after replacing the reference electrode;
rerun after replacing the Cl electrode;
monitor the temperature. If the ambient temperature is
too high, change the location of the system.
1007000700A5
ISE unit result error: Cl,
Na electrodes out of
slope range
Wrong electrode installation;
calibrator invalidated;
Electrodes performance
degrading.
air bubble in the electrode;
reference electrode damaged;
electrodes interact with each
other;
module or fluid path
temperature exceeding ℃;
Something wrong with the
sample
Reinstall the sensor;
Rerun after replacing the calibrator B;If the result is still
abnormal,, rerun after replacing calibrator A;
rerun after replacing problematic sensors.
Reinstall and recalibrate, after removing the bubble in
the electrode;
rerun after replacing the reference electrode;
rerun after replacing the Cl and Na electrode;
monitor the temperature. If the ambient temperature is
too high, change the location of the system.