3 Installation 3-11
Figure 3-7 Parameters Screen
The following table explains the parameters on the Parameters screen.
NOTE
Please set parameters according to instructions of reagents.
Improper settings may lead to unreliable test results.
Parameter Description
Test Name of the test.
No. No. of the test. It cannot be edited.
Full Name Full name of the test. It can be void.
Standard No. Standard No. of the test. It can be void.
Reac. Type Analyzing method, including Endpoint, Fixed-time and
Kinetic.
Pri. Wave. Primary wavelength to be used on the test.
Sec. Wave. Secondary wavelength to be used on the test. It can be void.
Direction It refers to the changing direction of the absorbance during
the reaction process. If the absorbance increases, select
Increase; otherwise, select Decrease.