14-4
14.5 Changing IBP Settings
14.5.1 Changing a Pressure for Monitoring
1. Select the pressure you want to change to enter its setup menu.
2. Select [Label] and then select your desired label from the list. The already displayed
labels cannot be selected.
Label Description Label Description
PA Pulmonary artery pressure CVP Central venous pressure
Ao Aortic pressure LAP Left atrial pressure
UAP Umbilical arterial pressure RAP Right atrial pressure
BAP Brachial arterial pressure ICP Intracranial pressure
FAP Femoral arterial pressure UVP Umbilical venous pressure
Art Arterial blood pressure P1 to P4 Non-specific pressure label
14.5.2 Choosing the Pressure Alarm Source
You can monitor for alarm conditions in systolic, diastolic and mean pressure, either singly
or in parallel. In the setup menu for the pressure, select [Alm Source] and choose from:
[Sys]: alarms are given only when the systolic pressure violates the alarm limits.
[Dia]: alarms are given only when the diastolic pressure violates the alarm limits.
[Mean]: alarms are given only when the mean pressure violates the alarm limits.
[Sys&Mean]: alarms are given when the systolic or mean pressure violates the alarm
limits.
[Mean&Dia]: alarms are given when the mean or diastolic pressure violates the alarm
limits.
[Sys&Dia]: alarms are given when the systolic or diastolic pressure violates the alarm
limits.
[All]: alarms are given when the systolic or mean or diastolic pressure violates the
alarm limits.