16-6 
 
For the microstream CO
2
 module, gas compensations are not required. 
 
For the maintream CO
2
 module, in the [CO
2
 Setup] menu, respectively select: 
  [Balance Gas] and toggle between [Room Air], [N
2
O] and [He]. Select [Room Air] 
when air predominates in the ventilation gas mixture, [N
2
O] when N
2
O predominates in 
the ventilation gas mixture, and [He] when He predominates in the ventilation gas 
mixture. 
  [O
2
 Compen] and then select [Off] or an appropriate setting according to the amount of 
O
2
 in the ventilation gas mixture. When the amount of O
2
 is less than 30%, you’d better 
switch this compensation off. 
  [AG Compen] and enter the concentration of anesthetic gas present in the ventilation 
gas mixture. This could compensate for the effect of AG on the readings. 
 
 
16.3.4 Setting up Humidity Compensation 
 
Sidestream and microstream CO
2
 modules are configured to compensate CO
2
 readings for 
either Body Temperature and Pressure, Saturated Gas (BTPS), to account for humidity in the 
patient’s breath, or Ambient Temperature and Pressure, Dry Gas (ATPD). 
1. ATPD: 
100/%)()(
22 ambco
PvolCOmmHgP
=
ambCO
PvolCOmmHgP
 
Where, 
2CO
P
= partial pressure, 
%vol
= CO
2
 concentration, 
amb
P
= ambient pressure, and 
unit is mmHg. 
 
As the mainstream CO
2
 module has a built-in heating component to prevent water vapor from 
condensing, setting humidity compensation is not needed. For the sidestream and 
microstream CO
2
 module, you can set the humidity compensation on or off according to the 
actual condition. To set the humidity compensation:: 
1. In the [CO
2
 Setup] menu, select [Humidity Compen]. 
2.  Select either [On] for BTPS or [Off] for ATPD, depending on which compensation 
applies.