16-10 
 
3.  After warm-up is finished, connect the sensor to a clean, dry airway adapter. The adapter 
should be vented to the air and isolated from CO
2
 sources, such as ventilator, the 
patient’s breathing, your own breathing, etc. 
4. Select [Start Zero Cal.] in the [CO
2
 Setup] menu. The message [CO
2
 Zero Running] 
is displayed. 
5.  It takes about 15 to 20 seconds. The message disappears when the zero calibration is 
completed. 
 
WARNING 
z  When perform a zero calibration during the measurement, disconnect the 
transducer from the patient’s airway first. 
 
 
16.9 Calibrating the Sensor 
For sidestream or microstream CO
2
 modules, a calibration should be performed once every 
year or when the readings go far beyond the range. For maintream CO
2
 module, no 
calibration is needed. For details, refer to the chapter Maintenance. 
 
 