21-3
21.3 Starting and Stopping Recordings
To manually start a recording, you can either:
Select the
hardkey on the front of either the patient monitor or the recorder module,
or
Select the [Record] button from the current menu or window.
Automatic recordings will be triggered in the following conditions:
Timed recordings will start automatically at preset intervals.
If both [Alarm] and [Alm Rec] for a measurement are set on, an alarm recording will be
triggered automatically as alarms occur.
To manually stop a recording, you can either:
Select the
hardkey again, or
Select [Clear All Tasks] in the [Record Setup] menu.
Recordings stop automatically when:
The runtime is over.
The recorder runs out of paper.
When the recorder has an alarm condition.
21.4 Setting up the Recorder
21.4.1 Accessing the Record Setup Menu
Select [Record Setup >>] in the main menu, and you can access the [Record Setup] menu.
21.4.2 Selecting Waveforms for Recording
The recorder can record up to 3 waveforms at a time. You can select, in turn, [Waveform 1],
[Waveform 2] and [Waveform 3] in the record setup menu, and then select the waveforms
you want. You can also turn off a waveform recording by selecting [Off]. These settings are
intended for realtime and scheduled recordings.