7-7 
 
7.5.2 Displaying Alarm Limits 
To have a better view of measurement numerics, you can choose not to display alarm limits 
in the parameter window by switching off displaying alarm limits in configuration mode. 
For details, refer to the chapter Managing Configuration. 
 
7.5.3 Setting Alarm Delay Time 
You can set the alarm delay time for over-limit alarms of continuously measured parameters. 
If the alarm triggered condition disappears within the delay time, the patient monitor will 
not issue the alarm. You can select [Alarm Setup>>] from [Main Menu] and then set 
[Alarm Delay]. 
 
In the configuration mode, you can follow these steps to set [Alarm Delay] and [ST Alarm 
Delay] respectively. 
 
1. Press the 
  hardkey on the monitor’s front to enter the main menu. 
2. Select [Manage Configuration >>] → [Configuration >>]. Enter the required 
password and then select [Ok]. 
3.  Select [View Config.]→ [Alarm] and set [Alarm Delay] and [ST Alarm Delay]. 
 
NOTE 
z  The alarm delay time can not be set in case of the following physiological alarms: 
Resp Apnea (The apnea time needs to be set independently), ST Alarm (The delay 
time needs to be set independently), Arrh. Alarm, PVCs Too High, ECG Weak 
Signal, 
Resp Artifact, SpO
2
 Desat, No pulse, Nellcor SpO
2
 over-limit alarm, FiO
2
 
Too Low, over-limit alarms of discontinuously measured parameters and MAC >3. 
 
 
7.5.4 Using Alarms 
Switching Individual Measurement Alarms On/Off 
1.  Select the parameter window for your desired measurement to enter its setup menu. 
2. Select [Alarm] and toggle between [On] and [Off]. 
  [On]: The patient monitor gives alarm indications in accordance with the preset 
alarm level and stores the related waveforms and parameters. 
  [Off]: The alarm off symbol “ ” is displayed in the measurement parameter 
window.