EasyManua.ls Logo

Mindray iPM-9800 - Removing Exhaust Gases from the System

Mindray iPM-9800
300 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
16-9
16.7 Removing Exhaust Gases from the System
WARNING
z Anesthetics: When using the Sidestream or Microstream CO
2
measurement on
patients who are receiving or have recently received anesthetics, connect the outlet
to a scavenging system, or to the anesthesia machine/ventilator, to avoid exposing
medical staff to anesthetics.
To remove the sample gas to a scavenging system, connect an exhaust tube to the gas outlet
connector of the module.
16.8 Zeroing the Sensor
The zero calibration eliminates the effect of baseline drift during CO
2
measurement exerted
on the readings and therefore maintains the accuracy of the CO
2
measurements.
16.8.1 For Sidestream and Microstream CO
2
Modules
For sidestream and microstream CO
2
modules, a zero calibration is carried out automatically
when necessary. You can also start a manual zero calibration if necessary. To manually start a
zero calibration, select [Start Zero Cal.] in the [CO
2
Setup] menu. Disconnecting the patient
airway is not required when performing a zero calibration.
16.8.2 For Mainstream CO
2
Modules
For mainstream CO
2
modules, zero the sensor whenever:
1. A new adapter is used;
2. You reconnect the sensor to the module;
3. You see the message [CO
2
Zero Required].In this case, check the airway adapter for
any blockage, e.g. mucus, etc. If a blockage is detected, clear or replace the adapter.
To zero the sensor, follow this procedure:
1. Connect the sensor to the module.
2. In the [CO
2
Setup] menu, set the [Operating Mode] to [Measure]. The message [CO
2
Sensor Warmup] is displayed.

Table of Contents

Other manuals for Mindray iPM-9800

Related product manuals