14-5 
 
14.5.3 Changing Averaging Time 
The IBP value displayed on the monitor screen is the average of data collected within a 
specific time. The shorter the averaging time is, the quicker the patient monitor responds to 
changes in the patient’s blood pressure. Contrarily, the longer the averaging time is, the 
slower the patient monitor responds to changes in the patient’s blood pressure, but the 
measurement accuracy will be improved. For critically ill patients, selecting shorter 
averaging time will help understanding the patient’s state. 
 
To set the averaging time, in the setup menu for the pressure, select [Response] and toggle 
between [High], [Med] and [Low]. The corresponding averaging time is about 1 s, 8 s and 
12 s respectively. 
 
 
14.5.4 Setting the Pressure Unit 
In the setup menu for the pressure, select [Unit] and toggle between [mmHg], [cmH
2
O] 
and [kPa]. 
 
 
14.5.5 Setting Wave Overlapping 
When two or more IBP channels are used, you can have IBP waves overlapped in the same 
area for a better view. 
In the waveform menu for the pressure, set [Wave Overlapping] as [On]. 
 
 
14.5.6 Setting Up the IBP Wave 
In the waveform menu for the pressure, you can: 
  Change the size of the pressure’s waveform by adjusting [Upper Scale], [Middle 
Scale] or [Lower Scale]. 
  When [Auto Scaling] is set on, the size of the pressure’s waveform will be adjusted 
automatically. 
 
When IBP waveforms are overlapped, you can select the overlapped waveforms to enter the 
[Overlap IBP Waves] menu. In the [Overlap IBP Waves] menu, you can: 
  In the [Arterial Pressure] and [Venous Pressure] area, adjust [Upper Scale], 
[Middle Scale] or [Lower Scale]. 
  In the [Select Wave] area, select the IBP waveforms you want to overlap.