8.8 12-Lead ECG Monitoring 
8.8.1 Entering the 12-lead ECG Monitoring Screen 
1.  Refer to the section 8.3.3 ECG Lead Placement for placing the electrodes. 
2.  Select the ECG parameter window and then set [Lead Set] to [12-Lead], and [ECG 
Display] to [12-Lead] in the popup menu. 
 
 
There are totally 12 ECG waves and 1 rhythm wave displayed on the screen. The rhythm 
lead is the HR-derived lead before entering the 12-lead ECG monitoring screen. The ST 
numerics are displayed in three groups: 
  ST Ant (anterior):  V1, V2, V3, V4 
  ST Inf (inferior):  II, III, aVF, (aVR) 
  ST Lat (lateral):  I, aVL, V5, V6 
Although aVR is displayed in the ST Inf group, it is not an inferior lead. 
 
Additionally, the 12-lead ECG monitoring has the following features: 
  The [Filter] mode is fixed to [Diagnostic] and cannot be changed. 
  The 
  hardkey on the monitor’s front is disabled.