A-1
A Product Specifications
A.1 Monitor Safety Specifications
A.1.1 Classifications
The patient monitor is classified, according to IEC60601-1:
Classification Name Specifications
Type of protection
against electric shock
Class I equipment, with both internal and external power supply
When you doubt the integrity of the external protective earth (ground)
or the protective earth conductor, switch the equipment to the internal
power supply (batteries).
Degree of protection
against electric shock
CO2/AG:
ECG/RESP/SpO2/NIBP/TEMP/IBP/CO:
BF (defibrillation proof)
CF (defibrillation proof)
Degree of protection
against hazards of
explosion
Not protected (ordinary)
Degree of protection
against harmful ingress
of water
In accordance with IPX1
Tested safety from vertically dripping water (DRIP-PROOF)
Mode of operation Continuous
Equipment type Portable