16-8 
 
16.4 Setting Barometric Pressure Compensation 
The mainstream CO
2 
module does not have the function of automatic barometric pressure 
compensation. For the mainstream CO
2 
module, the default barometric pressure is 760 mmHg. 
You must modify the barometric pressure based on the actual situation as follows: 
1.  In the main menu, select [Maintenance >>] → [User Maintenance >>] → enter the 
required password and then select [Maintain CO
2
] from the popup menu. 
2. Select [Barometric Pressure] and then enter the value of barometric pressure to which 
the patient monitor is exposed to. 
 
WARNING 
z  Be sure to set the altitude properly before using the mainstream CO
2
 module. 
Improper settings will result in erroneous CO
2
 reading. 
 
NOTE 
z  Both sidestream and microstream CO
2
 modules are configured with automatic 
barometric pressure compensation function. 
 
 
16.5 Measurement Limitations 
The following factors may influence the accuracy of measurement: 
  Leaks or internal venting of sampled gas 
  Mechanical shock 
  Cyclic pressure up to 10 kPa (100 cmH
2
O) 
  Other sources of interference, if any 
 
 
16.6 Troubleshooting the Sidestream CO
2
 Sampling 
System 
When the sampling system of the sidestream CO
2
 module works incorrectly, check if the 
sampling line is kinked. If not, remove it from the watertrap. If the monitor gives a message 
indicating the airway still works incorrectly, it indicates that the watertrap must have been 
blocked, and you should replace with a new one. Otherwise, you can determine that the 
sampling line must have been blocked. Replace with a new sampling line.