7-6
7.4.4 Switching off Alarm Sound
In configuration mode, you can set the minimum alarm volume. The minimum alarm
volume refers to the minimum value you can set for the alarm volume.
When the alarm volume is set to 0, the monitor stays in alarm sound off status and a
symbol appears on the screen.
The alarm sound off status is cancelled and the alarm volume returns to the factory default
value automatically when:
The monitor shuts down and then restarts.
The monitor changes to other alarm status, such as alarms paused status or alarms off
status.
The alarm volume restores the default value after the monitor reloads the
configuration.
WARNING
z When the alarm sound is switched off, the patient monitor will give no audible
alarm tones even if a new alarm occurs. Therefore the user should be very careful
about whether to switch off the alarm sound or not.
7.5 Alarm Configuration
7.5.1 Alarm Tone Configuration
Changing the Alarm Volume
1. In the main menu, select [Alarm Setup >>].
2. Select the appropriate volume from [Alm Volume]: X-10, in which X is the minimum
volume, depending on the set minimum alarm volume, and 10 the maximum volume.
Setting the Alarm Tone Interval
You can set the interval between alarm tones for high-level, medium-level and low-level
alarms. For details, refer to the chapter Managing Configuration.
WARNING
z Do not rely exclusively on the audible alarm system for patient monitoring.
Adjustment of alarm volume to a low level may result in a hazard to the patient.
Always keep the patient under close surveillance.