23-3
Transfer data within a LAN
Before transferring data within a LAN, do as follows:
1. Connect the patient monitor and the intended PC into the same LAN and acquire the
PC’s IP address.
2. Make sure that the data export system is active on the PC.
Follow the same procedure as via a crossover network cable to transfer data.
Transfer data via an external storage card
So far data can be transferred via CF storage card only.
1. Remove the external storage card from the patient monitor. For details, refer to the
Basic Operations section.
2. Run the data export system on the PC.
3. Insert the external storage card into the card reader that connects the PC.
4. Perform the data transfer operation following the help file of the system software.
23.4 Nurse Call
The patient monitor also provides nurse call signals to a nurse call system connected to the
monitor via the Auxiliary Output connector. To obtain nurse call signals, connect a nurse call
system to the monitor and then follow this procedure:
1. In the main menu, select [Maintenance >>]→[User Maintenance >>]→enter the
required password.
2. Select [Device Setup >>] to access the [Device Setup] menu.
3. Select [Auxiliary Output] and then [Nurse Call]. Then, you can change the following
settings:
Select [Signal Type] and toggle between [Pulse] and [Continuous].
[Pulse]: the nurse call signals are pulse signals and each pulse lasts 1 second. When
multiple alarms occur simultaneously, only one pulse signal is outputted. If an
alarm occurs but the previous one is not cleared yet, a new pulse signal will also be
outputted.
[Continuous]: the nurse call signal lasts until the alarm ends, i.e. the duration of a
nurse call signal equals to that of the alarm condition.