26-3
26.4 Calibrating ECG
The ECG signal may be inaccurate due to hardware or software problems. As a result, the
ECG wave amplitude becomes greater or smaller. In that case, you need to calibrate the ECG
module.
1. In the main menu, select [Maintenance >>][User Maintenance >>]enter the
required password and then select [Calibrate ECG] from the popup menu. A square
wave appears on the screen and the message [ECG Calibrating] is displayed in the
ECG waveform area.
2. Compare the amplitude of the square wave with the wave scale. The difference should
be within 5%.
3. After the calibration is completed, select [Stop Calibrating ECG].
You can print out the square wave and wave scale and then measure the difference between
them if necessary. If the difference exceeds 5%, contact your service personnel.
26.5 Calibrating the Touchscreen
1. In the main menu, select [Maintenance >>][User Maintenance >>]enter the
required password and then select [Cal. Touchscreen] from the popup menu.
2.
will, in turn, appear at different positions of the screen.
3. Select each
as it appears on the screen.
4. After the calibration is completed, the message [Screen Calibration Completed!] is
displayed. Select [Ok] to confirm the completion of the calibration.