7-10 
 
7.7 Clearing Technical Alarms 
For some technical alarms, their alarm lamp flashing and alarm tones are cleared and the 
alarm messages change to prompt messages after the 
  hardkey is pressed. After the 
patient monitor restores the normal alarm status, it can give alarm indications correctly in 
case these alarms are triggered again. 
 
For some technical alarms, all their alarm indications are cleared after the 
 hardkey is 
pressed. After the patient monitor restores the normal alarm status, it can give alarm 
indications correctly in case these alarms are triggered again. 
 
 
For others, their alarm tones are cleared but the alarm lamp flashing and alarm messages 
remain after the 
  hardkey is pressed. After the patient monitor restores the normal alarm 
status, all the alarm indications will continue if the alarm conditions still present. 
 
 
7.8 Testing Alarms 
When the monitor starts up, a self-test will be performed automatically. The system gives a 
beep indicating that the visible and audible alarm indicators are functioning correctly. In the 
meantime, the alarm lamp turns yellow then red. The start-up screen is displayed. 
 
For further testing of individual measurement alarms, perform the measurement on yourself 
(for example SpO
2
 or CO
2
) or using a simulator. Adjust alarm limits and check if alarms 
behave correctly. 
 
 
7.9 When an Alarm Occurs 
When an alarm occurs, observe the following steps and take proper actions: 
1.  Check the patient’s condition. 
2.  Confirm the alarming parameter or alarm category. 
3.  Identify the source of the alarm. 
4.  Take proper action to eliminate the alarm condition. 
5.  Make sure the alarm condition is corrected. 
 
For troubleshooting specific alarms, see appendix Alarm Messages.