8-20
8.7.6 Automatic Arrhythmia Relearn
Arrhythmia relearning is initiated automatically whenever:
The ECG lead or lead label is changed
The HR computing lead is changed.
The ECG lead is re-connected
A new patient is admitted
The paced status setting is changed
Arrhythmia analysis is switched on.
The ECG module becomes active.
After the calibration is completed, select [Stop Calibrating ECG]
8.7.7 Reviewing Arrhythmia Events
To review previously happened arrhythmia events, in the [ECG Setup] menu, select [Arrh.
Analysis >>]→[Arrh. Review >>].
In the [Arrh. Review] window, you can:
Select [Index] and then set a time as the index for finding your desired arrhythmia
events.
Rename an arrhythmia event by selecting the arrhythmia event you want to rename and
then selecting [Rename] to enter a new name for it.
Delete an arrhythmia event by selecting the one you want to delete and then selecting
[Delete Event].
Have more event selections by selecting
, , or .
After an event is selected, select [Waveforms] and then select
or beside
[Scroll] to navigate through the waveforms.
Print out the currently displayed arrhythmia waveforms and numerics through the
recorder by selecting [Record].
Print out the currently displayed arrhythmia waveforms and numerics through the
printer by selecting [Print].