7-4
NOTE
z When multiple alarms of different levels occur simultaneously, the patient monitor
will select the alarm of the highest level and give visual and audible alarm
indications accordingly.
7.3.5 Reminder Tones
When alarms are turned off or alarm tones are paused or turned off, the patient monitor will
give a single beep as the reminder tone in case of an active alarm condition. You can set the
reminder volume as [High], [Med] or [Low] in the [Alarm Setup] menu and set the
reminder interval as [1 min], [2 min] or [3 min] in configuration mode. You can only
switch on/off the reminder tone in configuration mode.
7.4 Understanding Alarm Statuses
You can set your monitor to one of the following four alarm statuses. The alarm symbols
match the alarm statuses as follows.
Alarms paused Alarms off
Alarm sound silenced Alarm sound off
7.4.1 Pausing Alarms
If you want to temporarily prevent alarms from sounding, you can pause alarms by pressing
the
hardkey on the monitor’s front.
No physiological alarm lamps flash and no alarms are sounded. No alarm messages are
shown.
No numeric and alarm limit flash.
The remaining pause time is displayed in the physiological alarm area.
For technical alarms, no alarms are sounded. The alarm lamps and alarm messages
remain active.
The
alarms paused symbol is displayed in the alarm status symbol area.