EasyManua.ls Logo

Mindray iPM-9800 - Stopping Report Printouts

Mindray iPM-9800
300 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
22-3
22.5 Stopping Report Printouts
To stop report printouts, select[Print Setup >>][Stop All Reports] in the main menu.
22.6 Setting Up Reports
22.6.1 Setting Up ECG Reports
You can print out ECG reports only under full-screen, half-screen or 12-lead monitoring
screen. To set up ECG reports, select [Print Setup >>][ECG Reports >>] in the main
menu.
[Amplitude]: set the amplitude of the ECG waveforms.
[Sweep]: set the wave print speed.
[Auto Interval]: If [Auto Interval] is set to [On], the system will automatically adjust
the space between waveforms to avoid overlapping.
[Gridlines]: choose whether to show gridlines.
[12-Lead Format]: If you select [12X1], 12 waveforms will be printed on a paper from
top to bottom. If you select [6X2], 12 waveforms will be printed from left to right with 6
waveforms on each side and a rhythm waveform will be printed at the bottommost.
22.6.2 Setting Up Tabular Trends Reports
To set up tabular trends reports, select [Print Setup >>][Tabular Trends Reports >>] in
the main menu.
Start time: You can set a time period whose trend data will be printed out by setting
[From] and [Back]. For example, if you set [From] as 2007-4-2 10:00:00 and [Back] as
[2 h], the outputted data will be from 2007-4-2 08:00:00 to 2007-4-2 10:00:00. In
addition, the [Back] can be set to either:
[Auto]: If [Report Layout] is set to [Time Oriented], the report will be printed by
time. If [Report Layout] is set to [Parameter Oriented], the report will be printed
by parameters.
[All]: If you select [All], all trend data will be printed out. In this case, it is no need
to set [From].
[Resolution]: choose the resolution of the tabular trends printed on the report.
[Report Layout]: If you select [Time Oriented], the report will be printed by time. If
you select [Parameter Oriented], the report will be printed by parameters.

Table of Contents