EasyManua.ls Logo

Mindray iPM-9800 - Calibrating CO2

Mindray iPM-9800
300 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
26-4
26.6 Calibrating CO2
For sidestream and microstream CO2 modules, a calibration is needed every year or when the
measured values have a great deviation. For maintream CO2 module, no calibration is
needed.
Tools required:
Gas bottle, with 4%, 5% or 6% of CO2.
T-shape connector
Tubing
Follow this procedure to perform a calibration:
1. Make sure that the CO2 module has been warmed up or started up.
2. In the main menu, select [Maintenance >>][User Maintenance >>]enter the
required password and then select [Maintain CO2 >>] from the popup menu.
3. In the [Maintain CO2] menu, select [Zero].
4. Connect the gas bottle with the tubing using a T-shape connector as shown below.
Check the airway and make sure there are no leaks.
5. Open the gas valve and vent CO2 into the tubing.
6. In the [Calibrate CO2] menu, enter the vented CO2 concentration in the [CO2] field.
7. In the [Maintain CO2] menu, the measured CO2 concentration, barometric pressure,
sensor temperature and current flowrate are displayed. After the measured CO2
concentration becomes stable, select [Calibrate CO2] to calibrate the CO2 module.
8. If the calibration is finished successfully, the message [Calibration Completed!] is
displayed. If the calibration failed, the message [Calibration Failed!] is displayed.
Perform another calibration.
9. Select [Exit] to exit the current menu.
Monitor
Gas bottle
Open to the air
Gas valve
Tubing

Table of Contents