20-6
20.4.3 Calculated Parameters
Abbreviation Unit Full spelling
PAO
2
mmHg partial pressure of oxygen in the alveoli
AaDO
2
mmHg alveolar-arterial oxygen difference
Pa/FiO
2
mmHg oxygenation ratio
a/AO
2
% arterial to alveolar oxygen ratio
MV L/min minute volume
Vd ml volume of physiological dead space
Vd/Vt % physiologic dead space in percent of tidal volume
VA L/min alveolar volume
20.5 Hemodynamic Calculations
20.5.1 Performing Calculations
1. In the main menu, select [Calc >>][Hemodynamic >>].
2. Enter values for calculation.
For a patient who is being monitored, [HR], [Art mean], [PA mean] and [CVP]
are automatically taken from the currently measured values. If you just have
performed C.O. measurements, [C.O.] is the average of multiple thermodilution
measurements. [Height] and [Weight] are the patient’s height and weight you
have entered. If the monitor does not provide these values, their fields appear
blank.
For a patient who is not being monitored, confirm the values you have entered.
3. Select the [Calculate] button. The system performs a calculation per the current
settings and displays the calculated values.
The values that exceed the range are displayed in yellow background. You can
select [Range] to view its normal range in the unit field. For those who are within
the range, their unit fields appear blank.
Invalid values are displayed as [---].
In the [Hemodynamic Calculation] window, you can:
Trigger a recording by selecting the [Record] button. The currently displayed renal
calculations are printed out by the recorder.
Review the previously performed calculations by selecting [Review].