15-3
3. Check if the height and weight are appropriate for your patient in the [Patient
Demographics] menu. Change if necessary.
4. Select the C.O. parameter window to enter the [C.O. Setup] menu. In the [C.O. Setup]
menu:
Check that the correct computation constant is entered. To change the
computation constant, select [Comp. Const] and then enter the correct value.
When a new catheter is used, the computation constant should be adjusted in
accordance with the manufacturer’s instructions for use.
Switch [Auto Ti] on and the injectate temperature is automatically obtained.
When [Auto Ti] is switched off, you can enter the injectate temperature in the
[Manual Ti] field.
5. Select [C.O. Measure] to enter the C.O. measurements window.
Patient monitor
C.O. cable
Injectate
Syringe
In-line probe
Balloon inflation port
PA catheter
Ice water
Hole
Stopcock