A-9
Pace Pulse
Pace pulses meeting the following conditions are labelled with a
PAC E marker:
Pace pulse markers
Amplitude:
Width:
Rise time:
±2 to ±700 mV
0.1 to 2 ms
10 to 100 µs
When tested in accordance with the ANSI/AAMI EC13:
Sections 4.1.4.1 and 4.1.4.3, the heart rate meter rejects all
pulses meeting the following conditions.
Pace pulse rejection
Amplitude:
Width:
Rise time:
Minimum input slew rate:
±2 to ±700 mV
0.1 to 2 ms
10 to 100 µs
50 V/s RTI
Alarm limit Range (bpm) Step (bpm)
HR High
Adult:
Pediatric, Neonate:
(low limit+2) to 300
(low limit+2) to 350
HR Low 15 to (high limit-2)
1
Basic algorithm
HR
3-, 5-, and 12-lead ECG
Neonate:
Pediatric:
Adult:
15 to 350 bpm
15 to 350 bpm
15 to 300 bpm
Measurement range
Interpretation of resting
12-lead ECG
Adult: 30 to 120 bpm
Resolution 1 bpm
Accuracy
3-, 5-, and 12-lead ECG: ±1 bpm or ±1%, whichever is greater.
Interpretation of resting 12-lead ECG: ±2 bpm
Sensitivity 200μV (lead II)