A-13
Sensor type Totally neonates Data Arms
518B 97 (51 male & 46 female) 200 pairs 2.38%
520N 122 (65 male & 57 female) 200 pairs 2.88%
The Pulse Oximeter with neonatal SpO2 sensors was also validated on adult subjects.
Refreshing rate 2 s
SpO
2
averaging time
7 s (When the sensitivity is set to High)
9 s (When the sensitivity is set to Medium)
11 s (When the sensitivity is set to Low)
Masimo SpO
2
Module
SpO
2
Standards Meet standards of ISO9919
Measurement range 1 to 100%
Resolution 1%
Accuracy
70 to 100%: ±2% (measured without motion in adult/pediatric
mode)
70 to 100%: ±3% (measured without motion in neonate mode)
70 to 100%: ±3% (measured with motion)
1% to 69%: Not specified.
Refreshing rate 1 s
SpO
2
averaging time 2-4 s, 4-6 s, 8 s, 10 s, 12 s, 14 s, 16 s
Low perfusion conditions
Pulse amplitude: >0.02%
Light penetration: >5%
Low perfusion SpO
2
accuracy ±2%