A-25
gas sample flow rate 120ml/min, using the DRYLINE™ watertrap and
neonatal DRYLINE™ sampling line (2.5m):
CO
2
≤250 ms (fall time: 200ms)
N
2
O ≤250 ms
O
2
≤600 ms
Hal, Iso, Sev, Des ≤300 ms
Enf ≤350 ms
gas sample flow rate 200ml/min, using the DRYLINE™ water trap and
adult DRYLINE™ sampling line (2.5m):
CO
2
≤250 ms (fall time: 200 ms)
N
2
O ≤250 ms
O
2
≤500 ms
Hal, Iso, Sev, Des ≤300 ms
Rise time
(10 % to 90%)
Enf ≤350 ms
Delay time <4 s
Primary anesthetic agent
In full accuracy mode: 0.15%,
In ISO accuracy mode: 0.4%
Anesthetic agent limit
(type AION 03)
Second anesthetic agent:
In full accuracy mode: 0.3% or 5% REL (10% in ISO accuracy mode) of
primary agent if primary agent is greater than 10%
In ISO accuracy mode: 0.5%
* Accuracy applies for the following conditions:
1. Measurements begin after the module warms up;
2. Ambient pressure is from 750 to 760 mmHg, and ambient temperature from 22 to
28ºC;
3. The measured gas is a dry gas and the balance gas N
2
;
4.Gas sample flow rate is 120 ml/min, respiration rate is 20 rpm with a fluctuation between
±3 rpm, and I:E is 1:2.