A-27
Alarm limit Range Step
EtCO2 High (low limit + 2) to 76 mmHg
EtCO2 Low 0 to (high limit - 2)mmHg
FiCO2 High (low limit + 2) to 76 mmHg
FiCO2 Low 0 to (high limit - 2)mmHg
1 mmHg
awRR High (low limit + 2) to 100 rpm
awRR Low 0 to (high limit - 2)rpm
1 rpm
EtO2 High (low limit + 0.3) to 100 %
EtO2 Low 18 to (high limit - 0.3)%
FiO2 High (low limit + 0.3) to 100 %
FiO2 Low 18 to (high limit - 0.3)%
0.1%
EtN
2
O High (low limit + 2) to 100 %
EtN
2
O Low 0 to (high limit - 2)%
FiN
2
O High (low limit + 2) to 100 %
FiN
2
O Low 0 to (high limit - 2)%
1%
EtHal/Enf/Iso High (low limit + 0.2) to 5.0 %
EtHal/Enf/Iso Low 0 to (high limit - 0.2)%
FiHal/Enf/Iso High (low limit + 0.2) to 5.0 %
FiHal/Enf/Iso Low 0 to (high limit - 0.2)%
0.1%
EtSev High (low limit + 0.2) to 8.0 %
EtSev Low 0 to (high limit - 0.2)%
FiSev High (low limit + 0.2) to 8.0 %
FiSev Low 0 to (high limit - 0.2)%
0.1%
EtDes High (low limit + 0.2) to 18.0 %
EtDes Low 0 to (high limit - 0.2)%
FiDes High (low limit + 0.2) to 18.0 %
FiDes Low 0 to (high limit - 0.2)%
0.1%