C-2
C.2 Parameters Configuration
C.2.1 ECG
ECG Setup
O.M Item Name
C M
General OR ICU NICU CCU User Defaults
Lead Set * * 5-Lead 3-Lead 5-Lead
ECG Waveforms * * 2 1 3
Alm Source * * HR
Alarm * * On
Alm Lev * * Med
Adu 120 140 120
Ped 160 180 160
HR High
Neo
**
200 200 200
Adu 50 40 50
Ped 75 60 75
HR Low
Neo
**
100 90 100
Sweep * * 25 mm/s
Beat Vol * * 2 1
Paced * No
Cascade * * Off
Notch Filter * On
HR Source * II
Gain * * X1
Filter * * Monitor Surgery Monitor Diagnostic
ECG Display * Normal
Smart Lead Off * * Off
Defib. Sync * Off
Pacemaker Rate * 60

Table of Contents