C-4
Arrh. Analysis
O.M Item Name Algorithm
C M
Genral OR ICU NICU CCU User Defaults
Arrh. Analysis / * * Off On
PVCs Alarm / * * Off On
Alm Lev / * * Med
PVCs High / * * 10 5 10
Arrhythmia Threshold Settings
Asystole * * 5 3 5
Vtac Rate * * 130 150 130
Vtac PVCs * * 6
Multif. PVC’s Window * * 15
Tachy * * Adu:
Ped:
Neo:
100
160
180
Brady
Mortara
* * Adu:
Ped:
Neo:
60
80
90
Arrhythmia Alarm Settings
X Alarm * * On
Asystole Alm Lev * * High
VFib/VTac Alm Lev * * High
X Alm Lev * * Med
X Alm Rec
Basic
* * Off
X Alarm * * On
Asystole Alm Lev * * High
VFib Alm Lev * * High
VTac Alm Lev * * High
X Alm Lev * * Med
X Alm Rec
Mortara
* * Off
All Alm Lev / * * Med
*X represents a certain arrhythmia event. Refer to the Specifications chapter for details. The
X in “X Alm Lev” refers to all arrhythmia events except for those specially marked ones.