System Diagnosis and Support 8-9
8.3.1 Turning on the System Configuration File is
Abnormal
8.3.2 The voltage of system power is abnormal
8.3.3 Temperature Alarming
Alarming tips
LOG record
Suggestion
Fail to open the file
"SystemConfiguration.ini", and
please check HDD data.
none
Reinstall the system software.
Alarming tips
LOG record
Suggestion
The real-time
battery will be
out of power,
please replace
quickly!
×××: System Monitor: Power supply alert! [VBAT], Current
voltage: [VVV] V, Limit voltage: [LLL]-[HHH] V.
×××represents time, e.g. 2010-12-25 14:15:25 [VVV]
represents the current value, and [LLL]-[HHH] represents the
lower and upper limits.
Replace the
button battery.
none
×××: System Monitor: Power supply alert! [XXX], Current
voltage: [VVV] V, Limit voltage:[LLL]-[HHH]V.
×××represents time, e.g. 2010-12-25 14:15:25
[XXX] represents voltage name, [VVV] represents the current
value, and [LLL]-[HHH] represents the lower and upper limits.
The voltage names respectively are:
AD+12V, A-12V, A-5V, AD+2V5, AD+1V8, D+1V5, D+1V2
Short-circuited,
protection and
overvoltage.
Check the
related circuit.
Alarming tips
LOG record
Suggestion
Temperature
alarming, XX
Shut down
(XX means
inversion timing,
starting 60S)
××× System Monitor: Temperature Alert! [XXX],
Current voltage: [VVV] degree, Limit temperature:
[LLL]-[HHH] degree
×××represents time, e.g. 2010-12-25 14:15:25
[XXX] represents voltage name, [VVV] represents the
current value, and [LLL]-[HHH] represents the lower
and upper limits. The detailed temperature name is
as following:
ADC thermal sensor, Multi FPGA thermal sensor,
Check the fan log
D:\z5\log (if the fan
stops working)/check
if the heat radiating
condition is normal.
The log records are the same
The temperature name is: power thermal sensor
as above
none
×××: System Monitor: Temperature Alert! [CPU
thermal sensor], Current voltage: [VVV] degree, Limit
temperature: [LLL] degree
×××represents time, e.g. 2010-12-25 14:15:25
The first temperature
of CPU
Suggestion as above.

Table of Contents

Other manuals for Mindray Z6

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Z6 and is the answer not in the manual?

Mindray Z6 Specifications

General IconGeneral
BrandMindray
ModelZ6
CategoryMedical Equipment
LanguageEnglish

Related product manuals