Care and Maintenance 9-9
Table 9-8 Mechanical Safety Check
a) Visually check to confirm there is no any crack.
b) Operate the casters to confirm the locking and
releasing functions are normal.
a) Visually check to confirm that there is no skewness
and the connecting screws are free of breakage or
falling off.
b) Check with the spanner to make sure that there is no
looseness between the caster and the base
connection screw.
1. Visually check to confirm there is no any crack.
2. Rock the handle gently and then lift the ultrasound
machine to confirm that the handle is free of
looseness and it can accept normal force.
Control panel
and turning
mechanism
Open the control panel to confirm that turning mechanism
is normal and the control panel is free of skewness and
looseness.
Display and
turning
mechanism
a) Visually check to confirm there is no any crack to the
cover of display.
b) Manually operate the monitor to make sure that the
turning angle of display is normal and no looseness
exists.
Other
mechanical
structures
Check to confirm that there is no looseness to other
mechanical parts, no crack to cover and no conductive
parts show in sight.
9.3.5 Electrical Safety Inspection
Only technical professionals from Mindray or engineers authorized by Mindray after training can
perform electric safety inspection.
Please refer to appendix A: Electrical Safety Inspection for details.